The Immunogold Silver Staining technique
(IGSS) finds application both at the electron microscope and the light
microscope level. In addition the technique is used to identify proteins or
nucleic acids after blotting.
Light microscopical and macroscopical
visualization of the latent gold signal requires an enhancement system that
renders a high contrast signal. Light insensitivity and negligible
autonucleation are required for ease of handling and low background. AURION
R-Gent SE-LM is a two-component reagent that meets these
requirements.
Introduction
The silver enhancement reaction is
based on the gold particle catalyzed reduction of Ag+ to metallic silver using
photographic developing compounds as the electron source. For light microscopy
and immunoblotting applications the generated silver signal should be of high
contrast. Furthermore the signal should be permanent and compatible with
counterstaining.
Product Description
The AURION R-GENT SE-LM components
constitute a Silver Enhancement Reagent which increases the average gold
cluster or particle size by deposition of metallic silver facilitating
detection at the light microscopical level. The generated brown-black signal is
also easily detected in bio assays and is compatible with counterstaining in
light microscopy. AURION R-GENT SE-LM has been tailored for the enhancement of
AURION Ultra Small Immunogold reagents and is equally suited for the
larger sized particles in the AURION Conventional Immunogold
reagents.
For enhancement, equal amounts of the DEVELOPER and ENHANCER
are mixed well and applied to the specimen. The enhancement mixture is
easy-to-use, exhibits extremely delayed auto-nucleation and can be used under
standard laboratory light conditions. Typical enhancement times are between 15
and 30 minutes. Auto-nucleation becomes visible only after 40-60 minutes. Light
microscopical specimens may be counterstained according to standard procedures.
The enhancement mixture has a pH value of 8.3-8.5.
AURION R-GENT SE-LM
is available as a kit in two unit sizes (2 x 30 ml or 2 x 250 ml) and consists
of a separate DEVELOPER and ENHANCER. The supplied amounts accomodate 600 and
5000 LM specimens respectively at 100 µl/specimen, or 60 and 500 bio
assay specimens at 1 ml/specimen. The reactivity is tested on dot-spots and the
absence of autonucleation is monitored by spectrophotometric techniques.
Light microscopy evaluation of tubulin
labeling with Ultra Small Immunogold Reagents and silver enhancement.
Applications
Detailed information is provided on
the package inserts.
Light Microscopy
Slide or Coverslip labeling
A few drops of the
freshly prepared enhancement mixture are applied to cover the specimen. During
enhancement the specimens are kept in a moist chamber. The ongoing enhancement
may at intervals be monitored with an inverted light microscope. When
enhancement is judged to be complete, the specimens are washed with distilled
water. A fixation step with photographic fixer is not required.
Whole mount
labeling
Specimens like floating (vibratome) sections can be enhanced
in Petri dishes or 6-24 well plates.
Bio Assays
Depending on the type of assay
the enhancement for bio assays can be performed in sealed plastic bags, Petri
dishes or in disposable screw cap sealed tubes.
Storage
The AURION R-GENT
SE-LM components are stored at 4°C and allowed to reach room temperature
before use.